Olympus FLUOVIEW FV1000 User Manual page 239

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Normal : Starts image acquisition when protocol comes the Track.
Trigger : Starts image acquisition after receiving external trigger input.
Wait
Condition: See the following table in case that [New mode] or [Series mode] is selected in
Sub-menu of
Condition column of New Mode
New mode
Change
XY
X
XYT
O
XYZ
X
XYZT
O
XT
O
XZ
X
XZT
O
Pt*T
X
Interval: When "Frame" is selected in column [Mode], time required to execute the track is
displayed. Change of [Interval] column cannot be done in case of [Series mode].
Frame Interval: Enter the interval time between image acquisitions in number of seconds. A
Image Size:
Zoom: Enter the zoom ratio in the range between 1.0 and 10.0.
: Starts image acquisition after wait time describes the track, the unit is
second.
[Mode].
When "Command" is selected in column [Mode], the macro command can be
enter to the track. For the macro commands detail, see section 2-4-2-2,
"Command List".
When "For" is selected in column [Mode], enter the repetition condition using a
variable. For the description of the repetition processing, see section 2-4-2-3 ,
"Protocol Repetition Processing".
Input details
Set 1 automatically
Set number of images to
be acquired
Set 1 automatically
Set number of images to
be acquired
Set number of lines
Set 1 automatically
Set number of lines
Set 1 automatically
number between 0.1 and 60000.0 can be entered.
Select acquisition image pixels. Select either [256 by 256], [320 by 240[,
[512 by 512], [640 by 640], [800 by 600], [1024 by 768], [1024 by 1024] or
[2048 by 2048] at [X-size by Y-size] list box.
APPLIED OPERATIONS
Condition column of the following
Series Mode
Enter number of slices to acquire
Enter number of slices to acquire
Enter number of lines to acquire
OPERATION INSTRUCTIONS
Page
/Protocol processor
2 - 1 5 5