Olympus FLUOVIEW FV1000 User Manual page 509

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

7.
Enter the value of the excitation wavelength into the [WaveLength 1] text box in the
[Excitation Wave Length] group box (e.g. 442).
8.
Enter the value of the emission wavelength into the [WaveLength 1] text box in the
[Emission Wave Length] group box (e.g. 480).
TIP
photometry, enter the value of the wavelength into the [WaveLength 2] text box
too.
9.
Click the <OK> button to close the [Edit Dye] dialog box.
10. Select the <Save New Setup> button to close the [FLUOVIEW Setup] dialog box.
TIP
Some dyeing method does not appear in [Available Dyes] list box in [Dyes]
subpanel.
It happens when there is no relation between the dyeing method and image
acquisition channel.
The relation of dye and the channel can be done by check [Assign dyes
manually] check box, or see section 1-5-4, "Associating a Detection Channel
and Filter to the Dyeing Method" for the setting.
Software Setup/
Adding the dyeing method
Page
MAINTENANCE
1 - 2 3