APPLIED OPERATIONS
<Start> button
OPERATION INSTRUCTIONS
Page
2 - 1 8 8
/ Protocol processor
The variable in the [For] statement can also be used by another cell in the protocol processor.
<Setting example>
It is also possible to change the laser intensity value at every image acquisition. In this case,
assume that the laser intensity of the Ar laser is variable N and create a protocol for acquiring
image as N varies from 10 to 100 in steps of 10.
1.
In Condition column to track No. 1, enter For statement "N = 10 To 100 Step 10".
2.
In track No. 2, specify the laser intensity value of the Ar laser as variable N.
3.
In track No. 3, specify the turning point of repetition processing.
4.
Click the <Start> button.
Acquire images by varying the laser intensity value of the Ar laser in 10% steps as
shown below
1st
Laser intensity %
One Point!
2nd
3rd
4th
10
20
30
5th
···
10th time
40
50
···
100