3.
In the [Mode] group box in the [Acquire] panel, click the [Surface XY-Norm] option
button, select <Sequen> from the list below it and select the scan mode to be used from
the displayed icons.
NOTE
TIP
4.
Using the <
use. This operation is not required if the current setting needs not be changed.
(Example: CH1 (FITC) and CH2 (PI) are set to Group 1.)
Click to increase the acquired group No.
Shows the group to be acquired.
Do not select <Line Aq> in the [Mode] group box in the [Acquire] panel
because the virtual channels are not compatible with Line Sequential
Scan.
If the observation mode is set to Sequential Scan, the [Dyes] sub-panel in the
[Acquire] panel cannot be used.
> and <
> buttons in [Group], set the group according to the reagent in
APPLIED OPERATIONS
Click to decrease the acquired group No.
OPERATION INSTRUCTIONS
Page
/Image Acquisition
2 - 8 3