Olympus FLUOVIEW FV1000 User Manual page 240

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

APPLIED OPERATIONS
OPERATION INSTRUCTIONS
Page
2 - 1 5 6
/ Protocol processor
Pan Status: Enter Pan value, refer to [Pan] [Zoom]group box in [Acquire] panel.
X Pan, Y Pan
FLUOVIEW software [Pan][Zoom] Group Box
TIP
Pan value can not enter the following condition.
· Mode is set to any of XY, XYT, XYZ or XYZT.
· SubMode is Normal.
· Zoom is set to 1 (1.0 x).
Variables are available in [For] statement for X Pan and Y Pan.
TIP
In addition, four-rule calculation is possible.
For details about how to use variables, see One Point of 2-4-8 "2 Example of
Setting Procedure of XYT Observation for hours (Protocol using repetition
processing)".
Angle: Assign rotated image scan angle of field of view. 0 through 359 can be entered. Angle
can not set at a certain sub scan mode, and a certain hardware. Please see 2-4-10
[Restrictions at [PAPP] window input time]. for entered angle value restriction.
Variables are available in [For] statement for Angle setting.
TIP
In addition, four-rule calculation is possible.
For details about how to use variables, see One Point of 2-4-8 "2 Example of
Setting Procedure of XYT Observation for hours (Protocol using repetition
processing)".