Olympus FLUOVIEW FV1000 User Manual page 492

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Software Setup/
Setting the System Configuration
MAINTENANCE
Page
1 - 6
3.
The dialog as shown below appears for use in saving the system configuration.
First, select the type of microscope and set the extra (intermediate attachment)
magnification to 1.0.
When the channel 3 of the FV1000 is selected, the [Fiber Port] check box will be
available. Please check here when a fiber port is connected to the FLUOVIEW.
Selecting "IX71" and "Automatic" in the [Microscope] group box or "BX61, 62", "BX61,
62WI", or "IX81" in the [Microscope] group box shows the [Microscope Setup] dialog
box to edit the BX operation panel (BX Control Panel). For details, see section 1-3-2
"Setting the [Microscope Control Panel]".
[Fiber Port] check box
[Excitation DM
Turret] check box
Check box appears
when selecting "2CH".
Check when using a
varialbe excitaion
dichroic mirror.