APPLIED OPERATIONS
The difference between the Frame and Line modes of Kalman Accumulation
In frame mode, the image is accumulated every one frame during acquisition.
In line mode, accumulation is done every one line during acquisition.
The frame mode is intended to utilize for fixed specimen observation, and the line mode is for
living specimen observation. Line mode makes it possible to shorten sampling interval so that it
is suitable for live cell observation.
OPERATION INSTRUCTIONS
Page
2 - 3 2
/Image Acquisition
3.
Enter the accumulation count in the text box and select the <Line> button displayed.
Enter the accumulation count.
TIP
The accumulation count can be set up a maximum of 64 times.
When 0 is set as the number of times of accumulation, the number is set to 1
automatically ..
4.
Click the <Once> button in the [Acquire] panel.
The acquired image will be displayed in the [Live] panel.
And the accumulated image can be displayed in the image window without being
displayed in full.
<Line> button