Olympus FLUOVIEW FV1000 User Manual page 28

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

Getting Started FLUOVIEW
OPERATION INSTRUCTIONS
Page
1-6
/Basic Operations
(4)
Condenser, polarizing plate
Condenser for transmitted lighting.
In addition, the rotary turret for the transmitted light DIC and the polarizing plate for
differential interference observation (analyzer) are also provided.
• To perform differential interference observation, engage the transmitted light DIC
(optional) matching the objective in use in the light path (For both visual observation
and laser differential interference observation).
• To perform visual differential interference observation or laser differential
interference observation, engage the polarizing plate in the light path.
(5)
Filters
These filters are used to adjust transmitted light.
• For transmitted observation using laser, disengage the LBD filter from the light path
and engage the FR (frost) filter in the light path by operating the filter levers. If the FR
filter is disengaged from the light path, the image may suffer from stripe interference.
(6)
Light path selector button
Select the light path between the visual observation and photography observation.
When <
> LED is lighted, visual observation can be done.
When <
> LED is lighted, TV or photography observation can be done.
(7)
Hand switch (U-FH is optionally available.)
This is the hand switch to operate the IX motorized system.