Getting Started FLUOVIEW
OPERATION INSTRUCTIONS
Page
1-2
/Basic Operations
(1)
Light path selector knob
Select the light path between the visual observation and photography observation.
See the following table and set the knob to the position corresponding to the
required light path.
Light path selector knob
Pushed in
Middle position
Pulled out
(2)
Cube turret
Select the fluorescence observation tube by rotating the turret.
Engage the desired cube in the light path for visual fluorescence observation or
visual transmitted light observation.
For laser microscopy, rotate the turret to page tab
state that mirror cube is entered.)
(3)
Analyzer IX2-MDICT
Polarizing plate for use in differential interference observation and polarized light
observation.
Rotate the turret to engage the IX2-MDICT in the light path for visual transmitted
light differential interference observation or transmitted polarized light observation.
(4)
Transmitted light DIC slider
SNPXLU of BX61WI)
This is the prism for use in differential interference observation.
Engage the transmitted light DIC in the light path for laser differential interference
observation or visual transmitted light differential interference observation.
Leaving the transmitted light DIC engaged during laser fluorescence observation will
degrade the resolution somewhat. We recommend disengaging the transmitted light
DIC from the light path when simple laser fluorescence observation is required.
Symbol
100% for the visual observation
20% for the visual observation, 80%
for photography observation
100% for photography observation
U-DICTS/WI-DICTHRA (built-in FV10-SRE or FV10-
Intensity Ratio
1
. (Set cube turret in the