Olympus FLUOVIEW FV1000 User Manual page 493

Confocal laser scanning biological microscope
Hide thumbs Also See for FLUOVIEW FV1000:
Table of Contents

Advertisement

4.
Click the [Z Stage] icon in [User Settings] to display the [Z Stage] panel in the front
position, and check and set the presence of motor controller, positioning of the motor
controller, backlash and moving amount of the Z motor.
[Z Stage] icon
TIP
The [Maximum range for Z series (mm)] text box accepts a setting value
between 1.0 and 3.0 (mm).
The values in the [Start Z] and [Stop Z] text boxes in the [Z Stage] sub-panel
are set according to the value set there.
For example:
When "1.0" (mm) is set as the maximum range for Z series, the minimum value
for Z starting becomes -500 (µm) and the maximum value for Z stopping
becomes 500 (µm).
Software Setup/
Setting the System Configuration
Page
Set the backlash of motor operation
in the [Z Stage] sub-panel.
Set the backlash during
scanning involving the Z stage.
Set the range where the stage
can move during the Z motor
operation.
Set the moving amount assigned to
the <Z stage coarse adjustment>
buttons.
Set the moving amount assigned
to the <Z stage fine adjustment>
buttons.
MAINTENANCE
1 - 7