Removing the Poppet Valves for the Replenishment Pumps

Use this procedure for removing the
POPPET VALVES in either the fixer or
the developer REPLENISHMENT
PUMP.
[1]
Deenergize the PROCESSOR. See page 1-5.
[2]
Lift the TOP COVER.
[3]
Remove the NON-DRIVE SIDE PANEL.
[4]
Use 2 CLAMPS TL-2170 on the 2 HOSES at
the REPLENISHMENT PUMP.
[5]
Loosen the HOSE CLAMPS and remove the
2 HOSES from the REPLENISHMENT PUMP.
[6]
Remove the 2 FITTINGS from the VALVE
BODY.
IMPORTANT
• Observe the direction of the POPPET
VALVES before you remove them.
• Removing the REPLENISHMENT PUMP
may be necessary for easier access to the
POPPET VALVES. See page 5-15 for the
removal procedure.
[7]
Remove the 2 O-RINGS and the 2 POPPET
VALVES from the VALVE BODY.
POPPET VALVE (2)
H104_0167BCA
H104_0167BA
Replacement of the Poppet Valves in the Replenishment Pumps
2B6845
NOTE
FITTING (2)
O-RING (2)
VALVE
BODY
[8]
Install all new O-RINGS and POPPET
VALVES in the VALVE BODY.
[9]
Check that the O-RINGS are seated correctly
and that the POPPET VALVES are in the
correct position in the VALVE BODY.
[10]
Reverse the above procedure to assemble.
Clamp TL-2170
developer
REPLENISHMENT
PUMP
fixer
REPLENISHMENT
PUMP
H104_0257AA
5-15