Replenishment Modes; Processor Configuration And Setpoints - Kodak X-Omat 270 Operator's Manual

Table of Contents

Advertisement

Processor
Configuration and
Setpoints
Item
*
Replenishment
Mode
*
Temperature
Lockout Mode
*
Display Units
*
Safelight
Receptacle Mode
*
Standby Mode
Developer
Temperature
Fixer Temperature
(minimum)
(Automatic)
Developer
Replenishment
(35 x 43 cm sheet)
(Automatic) Fixer
Replenishment
(35 x 43 cm sheet)
Transport Speed
Dryer Temperature
* Indicated parameters are not selectable on a cycle basis.
2B6841
Setup information consists of the film Processor configuration and process setpoints.
Setup information that was preset at the factory is listed in the table below.
Setup information is stored in battery backed-up RAM (Random Access Memory).
Therefore, you do not need to program new values every time that you turn on the
Processor. Even if power to the Processor is interrupted or the Processor is
turned off, you do not need to program the Processor.
If you wish, you may change the setup information by using the Soft Keys on the
Display Panel; See ''Processor Setup'' on page 22. The new setup information will be
retained in memory when power is interrupted or turned off.
Processor Default Configuration and Setpoints
K/RA
Automatic
Off
°F, in./min
Safelight
Interval
35.6°C
(96°F)
32.2°C
(90°F)
60 mL
85 mL
193 cm/min
(76 in./min)
48.9°C
(120°F)
RAPID
Automatic
Off
°F, in./min
Safelight
Interval
37.2°C
(99°F)
35°C
(95°F)
60 mL
85 mL
144.8 cm/min
106.7 cm/min
(57 in./min)
48.9°C
(120°F)
STANDARD
EXTENDED
Automatic
Off
°F, in./min
Safelight
Interval
34.4°C
(94°F)
32.2°C
(90°F)
60 mL
85 mL
53.3 cm/min
(42 in./min)
48.9°C
(120°F)
Automatic
Off
°F, in./min
Safelight
Interval
34.4°C
(94°F)
27.7°C
(85°F)
60 mL
85 mL
(21 in./min)
48.9°C
(120°F)
11