Adjusting the Detector Switches Up and Down
Before you can adjust the height of the
drive or non-drive side DETECTOR
SWITCH, both DETECTOR
SWITCHES must be moved to the
highest position.
[10]
On the drive side of the PROCESSOR:
(a)
Loosen the 2 SCREWS that hold the
DETECTOR SWITCH BRACKET to the
FILM DETECTOR HOUSING.
(b)
Move the DETECTOR SWITCH
BRACKET to the highest position.
[11]
Do Step [10] with the DETECTOR SWITCH
BRACKET on the non-drive side of the
PROCESSOR:
[12]
Connect the PORTABLE COMPUTER to the
15-PIN DIAGNOSTIC CONNECTOR on the
ELECTRICAL BOX.
See the User Instructions, Publication
Part No. 699614, for the operation of
the diagnostic procedures.
FILM
DETECTOR
HOUSING
DETECTOR
SWITCH
BRACKET
6-26
NOTE
NOTE
SCREWS
Adjusting the Height of the Detector Switch Bracket
[13]
Enter the ''SPECIFIC TEST MODE'' of the
diagnostic program on the PORTABLE
COMPUTER.
[14]
Select the ''PROCESSOR SENSOR TEST''.
The ''PROCESSOR SENSOR TEST'' will
provide you with the status of the 2 film
DETECTOR SWITCHES.
[15]
Adjust the drive side DETECTOR SWITCH.
(a)
Lift the top ENTRANCE DETECTOR
ROLLER, and insert the 2.5 cm (1 in.)
FILM STRIP between the ENTRANCE
DETECTOR ROLLERS on the drive
side of the PROCESSOR.
(b)
Slowly, move the DETECTOR
SWITCH BRACKET down until the
PORTABLE COMPUTER displays a
''Film detected'' message.
H104_0377BCB
H104_0377BA
2B6845