Checking the Replenishment Rates

DEVELOPER TUBING
METAL LATCH
H172_1005ACA
H172_1005AC
[5] Pull the DEVELOPER TUBING slightly and rotate it over the edge of the FRAME and into a GRADUATED
CYLINDER.
[6] Lift the UPPER ROLLER of the DETECTOR CROSSOVER AY for the correct time:
Cycle
Standard
Rapid
[7] When the REPLENISHMENT PUMP stops, check that the amount of solution in the GRADUATED
CYLINDER is correct for the film size and use condition. See the table on Page 32.
[8] If the replenishment rate is not correct, do the procedure "Adjusting the REPLENISHMENT PUMP" on Page
32.
[9] Connect the DEVELOPER TUBING by pushing it into the METAL LATCH until the DEVELOPER TUBING
snaps in place.
[10] Check the replenishment flow rate of the fixer solution by doing Steps 4 - 9 with the FIXER TUBING.
[11] Install the TOP COVER.
7C8772 – 12JAN99
QUICK
DISCONNECT
GRADUATED
CYLINDER
FIXER TUBING
Time
34 seconds
25 seconds
Adjustments and Replacements
UPPER ROLLER
DETECTOR CROSSOVER
ASSEMBLY
Warning
Wear protective eyewear. The replenishment solutions
are pumped quickly and might splash.
[1] Remove the TOP COVER.
[2] Lift the UPPER ROLLER of the DETECTOR
CROSSOVER AY.
[3] Check that the replenishment solutions flow
freely through the TUBING located along the
DRIVE SIDE RACKS.
[4] Disconnect the DEVELOPER TUBING by
pressing the METAL LATCH.
Note
The DEVELOPER TUBING might be identified with a
red WIRE TIE.
2 DETECTOR
SWITCHES
H172_1006BCA
H172_1006BC
31