Removing the Heat Exchanger

[1]
Deenergize the PROCESSOR. See page 1-5.
[2]
Lift the TOP COVER.
[3]
Remove:
• DRIVE SIDE PANEL
• WET SECTION COVER
• TOP EXIT RACK
• DRYER RACK
[4]
Use 2 CLAMPS TL-2170 on the 2 HOSES to
the HEAT EXCHANGER.
[5]
Remove the HOSE CLAMPS and the HOSES.
[6]
Remove and keep the 2 PLUGS, the
2 WASHERS, and the 2 O-RINGS from the
HEAT EXCHANGER. See the figure below.
[7]
Remove the HEAT EXCHANGER from the
WASH TANK.
Do not overtighten the PLUGS.
[8]
Reverse the above procedure to install a new
HEAT EXCHANGER and to assemble the
PROCESSOR.
5-8
Replacement of the Heat Exchanger
HEAT
EXCHANGER
IMPORTANT
The HEAT EXCHANGER should
extend from the TANK approximately
2.5 cm (1 in.). See the figure above.
HEAT EXCHANGER
2.54 cm
(1 in.)
H104_0198ACA
H104_0198AA
O-RING
(2)
WASHER
(2)
PLUG
(2)
H104_0208BCB
H104_0208BA
2B6845