Section 4: Replenishment Flow Rates
Figure 9
DEVELOPER TUBING
METAL LATCH
H172_1005ACA
H172_1005AC
[7] When the Replenishment Pump stops, check that the amount of solution in the Graduated Cylinder is correct
for the film size and use condition. See the table on Page 12. If the amount is not correct, contact qualified
service personnel.
[8] Connect the Developer Tubing again by pushing it into the Metal Latch until the Developer Tubing locks in place.
[9] Check the fixer replenishment flow rate by doing Steps
[10] Install the Top Cover.
Detector Crossover Assembly
Figure 10
7C8770 – 19JAN99
QUICK
Wear protective eyewear. The replenishment solutions
DISCONNECT
are pumped quickly and might splash.
GRADUATED
CYLINDER
FIXER TUBING
The Developer Tubing may have a red Wire Tie on it for
easy identification.
3
Warning
[1] Remove the Top Cover.
[2] Lift the Upper Roller of the Detector Crossover
Assembly.
[3] Check that the replenisher solutions flow freely
through the Tubing located along the drive side of
the Racks.
[4] Disconnect the Developer Tubing by pressing the
Metal Latch.
Note
[5] Pull the Developer Tubing slightly and rotate it
over the edge of the frame and into a Graduated
Cylinder.
[6] Lift the Upper Roller of the Detector Crossover
Assembly for the correct time:
Standard Cycle:
Rapid Cycle:
-
8
with the Fixer Tubing.
UPPER ROLLER
DETECTOR CROSSOVER
ASSEMBLY

Replenishment Flow Rates

34 seconds
25 seconds
2 DETECTOR
SWITCHES
H172_1006BCA
H172_1006BC
11