Setting Up the Processor

Calibrate the replenishment system whenever you add new chemicals to the
processor.
The operator calibrates the replenishment system for all process cycles in one
operation that determines the actual rate of solution flow through the replenishment
pumps. The operator will measure the volume of solution pumped during a set period
of time, then use the display panel to enter his measurement into the microprocessor.
The microprocessor computes the rate of solution flow through the pump, then adjusts
the period of time that the pump must operate to match the replenishment volume set
by the operator.
The volume actually measured during this procedure is not the volume
delivered for a 35x43 cm sheet of film.
[1]
To calibrate the replenishment system, press the ''GO TO SETUP'' key.
READY
135 F = DRYER SETPOINT
_______________________________________
|
/\
\/
|
DRYER
DRYER
|
[2]
Enter the 4 digit access code.
|
|
1
2
|
[3]
Press the ''OPTIONS'' key.
|
CYCLE
SETUP
|
PROC-
|
ESS
[4]
Press the ''MORE'' key.
|
REPLEN
TEMP
|
MODE
LOCK
|
NOTE
STD
|
|
|
SELECT
DISPLAY
GO TO
|
|
|
CYCLE
FIX
SETUP
|
|
|
TEMP
|
|
|
CANCEL
|
|
|
3
4
REQUEST
|
|
|
|
|
|
DONE/
|
|
|
OPTIONS
LANG
RETURN
|
|
|
|
|
|
DISPLAY
DONE/
|
|
|
UNITS
MORE
RETURN
|
|
|