Removing the Level Sensor Housing and the Level Sensor Probes

Use this procedure for either the red
developer or the blue fixer LEVEL
SENSOR PROBES.
[1]
Deenergize the PROCESSOR. See page 1-5.
[2]
Lift the TOP COVER.
[3]
Remove the WET SECTION COVER and the
NON-DRIVE SIDE PANEL.
[4]
Remove the 2 SCREWS and the WASHER
on the outside of the TANK.
[5]
Lift the LEVEL SENSOR HOUSING to remove
it.
[6]
Remove the LEVEL SENSOR PROBE from
the LEVEL SENSOR HOUSING.
[7]
Reverse the procedure to install new LEVEL
SENSOR PROBES or a new LEVEL
SENSOR HOUSING and to assemble the
PROCESSOR.
[8]
Check that the PROBE WIRE is correctly
attached to the SCREW and WASHER.
Install the GASKET SIDE of the WASHER
against the TANK. See the figure below.
GASKET SIDE
WASHER
SCREW
Correct Installation of the Washer
Against the Tank
2B6845
NOTE
TANK
PROBE
H127_0153ACA
WIRE
H127_0153AA
MANDREL
ASSEMBLY
SCREW (2)
WASHER
PROBE
WIRE
LEVEL SENSOR
HOUSING
Replacement of the Level Sensor Housing
and the Level Sensor Probes
REPLENISHER
HOSE
LEVEL
SENSOR
PROBE
H127_0134CCA
H127_0134CA
6-15