Adjusting the Detector Switches from Side to Side
[4]
Locate the DETECTOR SWITCH on the drive side of the PROCESSOR.
[5]
Loosen the 2 SCREWS that hold the DETECTOR SWITCH to the DETECTOR SWITCH BRACKET.
[6]
Move the DETECTOR SWITCH until the DETECTOR SWITCH is 4 - 6 mm (3/16 - 1/4 in.) from the edge
of the ACTUATOR.
[7]
Check that the DETECTOR SWITCH remains in the vertical position.
[8]
Tighten the 2 SCREWS that hold the DETECTOR SWITCH to the DETECTOR SWITCH BRACKET.
[9]
Do Steps [4] - [8] again for the DETECTOR SWITCH on the non-drive side of the PROCESSOR.
DETECTOR
SWITCH
BRACKET
2B6845
ACTUATOR
Adjusting the Detector Switch from Side to Side
SCREW
DETECTOR
SWITCH
4 mm-6 mm
(.248 in.-.372 in.)
DETECTOR
SWITCH
ACTUATOR
H127_0041BCC
H127_0041BA