Removing the Developer Filter

[1]
Deenergize the PROCESSOR. See page 1-5.
[2]
Lift the TOP COVER.
[3]
Remove:
• WET SECTION COVER
• REPLENISHMENT HOSE from the
MANDREL ASSEMBLY
[4]
Rotate the MANDREL ASSEMBLY until the
notch in the top of the ASSEMBLY aligns with
the LEVEL SENSOR HOUSING CLIP on the
LEVEL SENSOR HOUSING.
IMPORTANT
When removing the MANDREL
ASSEMBLY, place a DRIP TRAY
under it to prevent contamination of the
fixer.
[5]
Remove the MANDREL ASSEMBLY by
pulling straight up.
[6]
Remove:
• WING NUT from the MANDREL
ASSEMBLY
• DEVELOPER FILTER from the MANDREL
ASSEMBLY
[7]
Reverse the procedure to install a new
DEVELOPER FILTER.
[8]
Rotate the MANDREL ASSEMBLY until the
LEVEL SENSOR HOUSING CLIP is in the
locked position. See the figure below.
H127_0061BCA
H127_0061BA
2B6845
MANDREL
ASSEMBLY
DEVELOPER
FILTER
WING NUT
LEVEL
SENSOR
HOUSING
LEVEL SENSOR
HOUSING CLIP
MANDREL
UNLOCKED
HOUSING
REPLENISHMENT
HOSE
H127_0135CCB
H127_0135CA
LOCKED
5-5