Replenishment; Checking The Replenishment Flow Rates, If The Processor Has Been Set For "High Volume; Adjusting The Replenishment Pump - Kodak M35M X-Omat Operator's Manual

Table of Contents

Advertisement

OPERATOR MANUAL
Section 4: Replenishment
Checking the Replenishment Flow Rates
Checking the Replenishment Flow Rates
Figure 12
DEVELOPER TUBING
METAL
LATCH
H112_0114ACA
H112_0114AC
[7] When the Replenishment Pump stops, check that the amount of solution in the Graduated Cylinder is the same
as in the table on Page 13. If the amount is not correct, contact qualified service personnel.
[8] Connect the Developer Tubing again by pushing it into the Metal Latch until the Developer Tubing locks in
place.
[9] Check the fixer replenishment flow rate by doing Steps 3 - 8 with the Fixer Tubing.
[10] Install the Top Cover.
Detector Crossover Assembly
Figure 13
12
QUICK
DISCONNECT
GRADUATED
CYLINDER
FIXER TUBING
Use only qualified service personnel to perform
all necessary adjustments and procedures.
[1] Remove the Top Cover.
[2] Lift the Upper Roller of the Detector Crossover
Assembly. See Figure 13.
[3] Check that the replenisher solutions flow freely
through the tubing located along the drive side of
the Racks.
[4] Disconnect the Developer Tubing by pressing the
Metal Latch.
Note
The Developer Tubing may have a red wire tie on it for
easy identification.
[5] Pull the Developer Tubing slightly and rotate it
over the edge of the frame and into a Graduated
Cylinder.
[6] Lift the Upper Roller of the Detector Crossover
Assembly for the correct time. See the table on
the next page.
UPPER ROLLER
DETECTOR
CROSSOVER
ASSEMBLY
DETECTOR
SWITCH (2)
H112_0121BCB
H112_0121BA
09JUL98 – 981799