Hide thumbs Also See for X-Omat 2000:

Advertisement

Quick Links

{OperatorsManual}{Production}{HealthImaging}
Publication No. 7C8770
19JAN99
Supersedes 7C8770
03DEC98
HEALTH IMAGING
OPERATOR MANUAL
for the
Kodak X-Omat 2000 PROCESSOR
Kodak X-Omat 2000A PROCESSOR
© Eastman Kodak Company, 1999

Advertisement

Table of Contents
loading

Summary of Contents for Kodak X-Omat 2000

  • Page 1 {OperatorsManual}{Production}{HealthImaging} Publication No. 7C8770 19JAN99 Supersedes 7C8770 03DEC98 HEALTH IMAGING OPERATOR MANUAL for the Kodak X-Omat 2000 PROCESSOR Kodak X-Omat 2000A PROCESSOR © Eastman Kodak Company, 1999...
  • Page 2: Table Of Contents

    X-Omat Processor, call a plumber or other contractor to correct any problem with the drain. Kodak accepts no responsibility or liability whatsoever for the serviceability of any drain connected to or associated with a Kodak X-Omat Processor. Such drains are the sole responsibility of the customer.
  • Page 3: Overview

    Overview Section 1: Overview The Kodak X-Omat 2000 and 2000A Processors are designed to process medical x-ray sheet or roll film that can be processed in Kodak RP X-Omat Developer and Replenisher and in Kodak RP X-Omat Fixer and Replenisher.
  • Page 4: Processing Film

    OPERATORS MANUAL Section 2: Processing Film Start-Up Main Power Switch Figure 1 RUN/STANDBY Warning TOP COVER SWITCH Keep fingers, clothing, cleaning materials, etc., away from the moving parts of the Processor. [1] Turn on the water supply and the Main Power Switch.
  • Page 5 [7] Press the Run/Standby Switch. WASH RACK [8] When the Temperature Ready Light flashes, feed a sheet of Kodak Roller Transport Cleanup Film into the Processor. [9] Adjust the Dryer Temperature Control Knob to the lowest setting to dry the film.
  • Page 6: Feeding Sheet Film

    OPERATORS MANUAL Feeding Sheet Film Film Sizes and Direction of Film Feed Figure 4 [1] Press the Run/Standby Switch. Note 1 each 1 each • The Motors will begin to operate. 35 x 35 cm • After film removal, the Processor will automatically return to the standby mode.
  • Page 7: Shutdown

    Processing Film Shutdown Processor in Shutdown Mode Figure 5 [1] Turn off the Main Power Switch and the water supply. TOP COVER [2] Disconnect the main power. [3] Remove the Top Cover and Evaporation Covers. [4] Do the “Daily procedure” on Page 15. Important Keep the Top Cover open overnight for venting purposes.
  • Page 8: Processing Solutions

    OPERATORS MANUAL Section 3: Processing Solutions Preparing the Solutions Processing Tanks with the Splash Guard Figure 6 and the Drip Tray Caution DRIP TRAY • Avoid contamination of the processing solutions. SPLASH GUARD • To avoid spilling solutions from one Tank into the next one, use the Splash Guard between the developer and fixer Tanks.
  • Page 9 Processing Solutions Draining the Tanks Figure 7 WEIRS WATER INLET WASH WASH FIXER FIXER 3 DRAINS DEVELOPER DEVELOPER H172_0503DCA H172_0503DC [1] If the Processor has a Silver Recovery Unit connected, disconnect the Fixer Drain Tube at the input side of the Silver Recovery Unit.
  • Page 10: Filling The Tanks

    (a) Fill the developer Tank half full of developer replenisher from the replenisher tank. (b) Add 190 mL (6.5 fl oz) of Kodak RP X-Omat Developer Starter. (c) Fill the developer Tank to the higher Fill Line on the red Weir with developer replenisher.
  • Page 11: Replenishment Flow Rates

    Replenishment Flow Rates Section 4: Replenishment Flow Rates Figure 9 DEVELOPER TUBING Warning QUICK Wear protective eyewear. The replenishment solutions DISCONNECT METAL LATCH are pumped quickly and might splash. GRADUATED [1] Remove the Top Cover. CYLINDER [2] Lift the Upper Roller of the Detector Crossover FIXER TUBING Assembly.
  • Page 12 Medium 25 to 75 sheets 25 sheets or less* * If sensitometry does not stay within control limits, flooded replenishment may be needed. Contact your Kodak Representative to order and install the Flooded Replenishment Kit, Part No. 419779. Note Refer to Service Bulletin 30, Part No. 632661, for additional information and the latest processing recommendations.
  • Page 13: Correcting Difficulties

    Correcting Difficulties Section 5: Correcting Difficulties 1. Transport Failure 2. Surface Artifacts 3. Abnormal Film Densities 4. Wet Films 5. Low Solution Levels 6. Overlapping of Films • • Film Feeding Error Feed only single thicknesses of film. Feed the next film only after the film feed signal sounds If there is no film feed signal, refer the difficulty to qualified service personnel.
  • Page 14 OPERATORS MANUAL 1. Transport Failure 2. Surface Artifacts 3. Abnormal Film Densities 4. Wet Films 5. Low Solution Levels 6. Overlapping of Films • Check that the correct Bulb and Safelight Filter are in the Safelight and at the correct distance from the Feed Shelf and work surface. •...
  • Page 15: Cleanup And Periodic Maintenance

    Cleanup and Periodic Maintenance Section 6: Cleanup and Periodic Maintenance Daily Procedure The best time to do daily cleanup is at shutdown. Daily Cleaning of the Racks and Figure 11 Crossover Assemblies Warning • Dangerous Voltage. Disconnect the main power before doing any of the following cleanup procedures.
  • Page 16: Every Two Weeks

    OPERATORS MANUAL Every Two Weeks Biweekly Cleaning of the Racks and Figure 12 Crossover Assemblies Warning Dangerous Voltage. Disconnect the main power before doing any of the following cleanup procedures. [1] Turn off the Main Power Switch. [2] Disconnect the main power. [3] Turn off the water supply.
  • Page 17 SIDE PATES If solution is present in the Bypass, the Silver Recovery Unit is not operating. [16] Use Kodak Silver Estimating Test Papers to check the silver content of the solution flowing into the drain. [17] If the test shows more than 1 gram/liter of silver, install a new cartridge in the Silver Recovery Unit.
  • Page 18: Special Maintenance For Low Or High Volume Usage

    [4] To fill the fixer Tank, add fixer replenisher until the solution is at the higher Fill Line on the blue Weir. [5] To fill the processor developer Tank: (a) Add 190 mL (6.5 fl oz) of Kodak RP X-Omat Developer Starter. (b) Fill the developer Tank to the higher Fill Line on the red Weir with developer-replenisher solution.
  • Page 19 Cleanup and Periodic Maintenance Splash Guard and Drip Tray Figure 16 DRIP TRAY SPLASH GUARD H172_1003ACA H172_1003AC Caution When installing the Racks, use the Splash Guard. Lower the Racks slowly. [6] Install the Racks, the Crossover Assemblies, and the Dryer Assembly. Check that each Assembly seats firmly. [7] Turn on the main power, the water, and the Processor.
  • Page 20: New Equipment Warranty

    Kodak. – Such damages for which Kodak will not be responsible, include, but are not limited to, loss of revenue or profit, downtime costs, loss of use of the equipment, cost of any substitute equipment, facilities or services or claims of your customers for such damages.

This manual is also suitable for:

X-omat 2000a

Table of Contents