Removing the Developer Thermistor RT1

[1]
Deenergize the PROCESSOR. See page 1-5.
[2]
Lift the TOP COVER.
[3]
Remove the DRIVE SIDE PANEL.
[4]
Drain the DEVELOPER TANK until the
developer is below the RECIRCULATION
OUTPUT opening.
[5]
Use a CLAMP TL-2170 on the longer HOSE
to the DEVELOPER THERMOWELL.
[6]
Disconnect the CONNECTOR P/J 26.
[7]
Remove the DEVELOPER THERMISTOR
RT1 from the DEVELOPER THERMOWELL.
• Use SEALANT TL-3230 on the
DEVELOPER THERMISTOR when
installing the DEVELOPER
THERMISTOR in the DEVELOPER
THERMOWELL. Use only TL-3230.
Any other SEALANT will corrode the
plastic in the THERMOWELL.
• Do not overtighten the parts when installing
the DEVELOPER TERMISTOR.
[8]
Reverse the above procedure to install a new
DEVELOPER THERMISTOR.
[9]
Check for leakage at the DEVELOPER
THERMOWELL.