Ametek Elgar ReFlex Power Operation Manual page 358

Table of Contents

Advertisement

ReFlex Power™
8. Set the second power voltage to 10.0 Volts.
9. After the DMM voltage stabilizes enter the voltage of the DMM
Note: This voltage should be equal to 10.0 Volts.
10. Send the command to calculate the gain and offset of the IAI board.
11. Set the active load back to the SCPI mode:
12. Send the system error query command to make sure calibration was
13. If there is no error returned at the system error query command then
14. Reset the unit
M380056-01 Rev L
(Connected to second power supply. or Power supply connected to
IAI Board)
CAL<n>:INP:VOLT2 <value-2>
CAL<n>:MEAS:OFFSET:CALC
INP<n>:CURRent:EXTernal 0
successful.
SYST:ERR?
save the calibration result with the following command
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
Active Loads
7-83