ReFlex Power™
D
4.5.4
EFAULT
P
OWER
When an AC power supply module is first powered on, the following
parameters are set to factory default values:
R
ESET
If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:
M380056-01 Rev L
O
PERATIONAL
-O
C
N
ONDITIONS
Output configuration: standalone
Output voltage: 0V
Output current: full scale
Overvoltage protection (OVP): 106% of full scale voltage; programmable
mode
Undervoltage protection (UVP): 0V, programmable mode
Overcurrent protection (OCP): 106% of full scale current; programmable
mode
Output state: off
Output isolation and remote sense relays: open
Frequency set to 60 Hz
C
ONDITIONS
Output configuration: unchanged
Output voltage: 0V
Output current: full scale
Overvoltage protection (OVP): 106% of full scale voltage; programmable
mode
Undervoltage protection (UVP): 0V, programmable mode
Overcurrent protection (OCP): 106% of full scale current; programmable
mode
Output state: off
Output isolation and remote sense relays: open
Frequency set to 60 Hz
All status reporting data structures are cleared. Enable-masks are not
cleared.
C
ONDITIONS
AC Power Supplies
4-39