ReFlex Power™
4. Start the calibration of the gain pot
5. Set the load box to approximately 6.5Amps. Wait for the current to stabilize then
make the AC current measurement using the meter connected to the power
supplies output lead.
6. Finish the gain pot calibration
7. Send the system error query command to make sure calibration was successful.
8. If there is no error returned at the system error query command then save the
calibration result with the following commands
9. Optional Commands:
10. Reset the unit
M380056-01 Rev L
CAL<n>:CALPOTS:GAIN:START
CAL<n>:CALPOTS:GAIN:CURR <value>
CAL<n>:CALPOTS:GAIN:CALC
SYST:ERR?
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
T
o query the offset pot set value
To query the gain pot set value: CAL<n>:CALPOTS:GAIN?
*RST<n>
:
CAL<n>:CALPOTS:OFFSET?
AC Power Supplies
4-47