Active Loads
7. Set the DAC count to 50.0% of its full range
8. Wait for the DC current to stabilize. After the current reading is
9. Set the DAC count to 72.5% of its full range
10. Wait for the DC current to stabilize. After the current reading is
11. Set the DAC count to 95% of its full range
12. Wait for the DC current to stabilize. After the current reading is
13. Set the DAC to zero counts
7-58
CAL<n>:INP:CURR:COUNTS 32000
stabilized, enter the current using following command
CAL<n>:INP:CURR:FIVEPOINT3 <value-3>
CAL<n>:INP:CURR:COUNTS 43000
stabilized, enter the current using following command
CAL<n>:INP:CURR:FIVEPOINT4 <value-4>
CAL<n>:INP:CURR:COUNTS 56000
stabilized, enter the current using following command
CAL<n>:INP:CURR:FIVEPOINT5 <fval>
CAL<n>:INP:CURR:COUNTS 0
ReFlex Power™
M380056-01 Rev L