Calibration; Scope; Recommended Calibration Equipment; Calibration Setup - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

Fixed Power DC Power Supply

CALIBRATION

5.6
Calibration is not normally required after the FPDC is shipped. However, if
necessary, calibration is easily performed through the Controller module interface
with SCPI commands, requiring only a meter and a mating connector.
S
5.6.1
COPE
The factory calibration consists of setting the 4 outputs to nominal, that is +5VDC
= 5.000, etc. The following procedures can be used to reestablish this as
required.
R
5.6.2
ECOMMENDED
DMM, with at least 6 ½ digit accuracy and must be within its calibration period.
Calibration Power Connector:
C
5.6.3
ALIBRATION
Calibration can be done without load, therefore only a meter and a specially
configured mating connector is required, made up as follows:
(Note: these Test Connectors are used during performance verification as well).
FPDC P
OWER
In all cases, there is a wire for each of the 4 outputs and their corresponding
returns. These should be two to three inches in length and have the bare ends
soldered, (tinned) for convenient connection to the DMM. These are labeled as
Stub in the Type column.
5-28
C
ALIBRATION
Mating connector: 21 pin D_SUB, special: Positronic Subminiature-D
CBC Series, CBC21WA4M1000Z (crimp signal contacts)
Four male contacts; high-current male contact ratings: MC4012D rated at
20A with 12AWG wire
22 AWG insulated wires
S
ETUP
C
C
ONNECTOR FOR
E
QUIPMENT
ALIBRATION
ReFlex Power™
M380056-01 Rev L