Cal Pot Calibration For Voltage Range Low; Figure 4 14. Calibration Instrumentation Setup - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

AC Power Supplies
C
P
4.6.4
AL
The following procedure calibrates cal-pot for the voltage range low. This
calibration is done with the AC current meter connected to the output lead of the
AC power supply.
1. Connect the AC current meter to the output lead of the power supply. Ensure that
no load is connected to the output.
2. Issue clear and reset commands:
3. Start the internal calibration of the offset pot
4-46
Figure 4-14. Calibration Instrumentation Setup
C
OT
ALIBRATION FOR
*CLS
*RST<n>
CAL<n>:CALPOTS:OFFSET:CALC
V
R
OLTAGE
ANGE
ReFlex Power™
L
OW
M380056-01 Rev L