ReFlex Power™
14. Set the output frequency to 250Hz
15. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
16. Set the output frequency to 440Hz
17. Wait for AC current to stabilize. Measure the output current using the AC current
meter
18. Set the output frequency to 630Hz:
19. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
20. Set the output frequency to 820Hz:
21. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
22. Set the output frequency to 1010Hz
23. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
24. Set the output frequency to 1200Hz:
25. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
M380056-01 Rev L
SOUR<n>:FREQ 250
CAL<n>:OUTP:FREQ:CURR2 <value-2>
SOUR<n>:FREQ 440
CAL<n>:OUTP:FREQ:CURR3 <value-3>
SOUR<n>:FREQ 630
CAL<n>:OUTP:FREQ:CURR4 <value-4>
SOUR<n>:FREQ 820
CAL<n>:OUTP:FREQ:CURR5 <value-5>
SOUR<n>:FREQ 1010
CAL<n>:OUTP:FREQ:CURR6 <value-6>
SOUR<n>:FREQ 1200
AC Power Supplies
4-83