Active Loads
LPAL V
7.6.23
This procedure calibrates the voltage for voltage gain 4 and current range
high.
between the positive and negative lead of the active load. Also connect the
sense positive-to-positive lead and sense negative-to-negative lead.
This procedure requires maximum voltage 109 V and maximum current 3.0
Amps.
1. Turn the active load on
2. Set the conductance count to 0x6666
3. Set the input DC voltage to 27 Volts. After the input DC voltage
4. Set the input DC voltage to 109 Volts. After the input DC voltage
5. Send the calibration command to calculate the gain and offset:
6. Send the system error query command to make sure calibration was
7-64
C
OLTAGE
ALIBRATION
Make sure that the shunt and the DC voltmeter are connected
INP<n>:STATE 1
CAL<n>:INP:COND:COUNTS 0x6666
stabilizes enter the first exact voltage using following command
CAL<n>:INP:VOLT 1 <value-1>
stabilizes enter the second exact voltage using following command
CAL<n>:INP:VOLT 2 <value-2>
CAL<n>:MEAS:OFFSET:CALC
successful.
SYST:ERR?
– P
II
V
ART
OF
ReFlex Power™
M380056-01 Rev L