Hpal Voltage Calibration - Part Iv Ofv - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

ReFlex Power™
HPAL V
7.6.11
This procedure calibrates the voltage for voltage gain 1 and current range high.
Make sure that shunt and the DC voltmeter is connected between the positive
and negative lead of the active load. Also connect the sense positive-to-positive
lead and sense negative-to-negative lead.
This procedure requires maximum voltage 500.0V and maximum current 1.0
Amps
1. Turn the active load on:
2. Set the conductance count to 0x831:
3. Set the input DC voltage to 10 Volts. After the input DC voltage
4. Set the input DC voltage to 500 Volts. After the input DC voltage
5. Send the calibration command to calculate the gain and offset:
6. Send the system error query command to make sure calibration was
7. If there is no error returned at the system error query command then
8. Reset the unit
M380056-01 Rev L
C
OLTAGE
ALIBRATION
INP<n>:STATE 1
CAL<n>:INP:COND:COUNTS 0x831
stabilizes enter the first exact voltage:
CAL<n>:INP:VOLT 1 <value-1>
stabilizes enter the second exact voltage:
CAL<n>:INP:VOLT 2 <value-2>
CAL<n>:MEAS:OFFSET:CALC
successful:
SYST:ERR?
save the calibration result as follows:
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
– P
IV
V
ART
OF
Active Loads
7-41