12. Wait for the DC current to stabilize. After the current reading is stabilized,
13. Set DAC counts to zero:
14. Send calibration command to make sure that all the DAC counts and
15. Send the system error query command to make sure calibration was
16. If there is no error returned at the system error query command then save
17. Reset the unit:
M380056-01 Rev L
enter the current using following command:
CAL<n>:INP:CURR:FIVEPOINT5 <fval>
CAL12:INP:CURR:COUNTS 0
current values are valid:
CAL<n>:INP:CURR:CALC
successful:
SYST:ERR?
the calibration result with the following commands:
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
Active Loads