AC Power Supplies
26. Perform the voltage calibration using following command
27. Send the system error query command to make sure calibration was successful.
28. If there is no error returned at the system error query command then save the
calibration result with the following commands
29. Reset the unit
4-78
CAL<n>:OUTP:FREQ:CALC
SYST:ERR?
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
ReFlex Power™
M380056-01 Rev L