Calibration; Scope - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

AC Power Supplies
6. Turn on the isolation and remote sense relays, turn on the output
7. Measure the output voltage and current with the external instruments.
8. Ensure that the actual values of the output parameters are within
9. Query the values of the output parameters with the following commands:
10. Ensure that the readback values of the output parameters are within
11. Turn off the output with the following command:
4.6

CALIBRATION

S
4.6.1
COPE
Procedures are provided in the following sections for calibration of the ReFlex
Power
RFP-A301K-875-XXXX, and ReFlex Power
A305K-875-XXXX. Calibration is easily performed through the Controller module
interface with SCPI commands, requiring only 2 meters and resistive load.
4-44
converters, and program full scale voltage and current with the following
commands. The parameter, <fval>, is the full scale output voltage or
current of the particular model under test. The parameter, [1000+n1], is
the sum of 1000 and the slot number of the master module. For example,
if the master module is located in slot 4, the parameter value would be
1001.
OUTP[1000+n1]:STATE 1
SOUR[1000+n1]:CURR <fval>
SOUR[1000+n1]:VOLT <fval>
specifications of the programmed values. The default frequency is 60 Hz.
MEAS[1000+n1]:VOLT?
MEAS[1000+n1]:CURR?
MEAS[1000+N1]:FREQ?
specifications of the actual values.
*RST[n]
TM
AC Power supply module: ReFlex Power
TM
ACPS 1.2KHz module, P/N:
TM
ACPS 5.0KHz module, P/N: RFP-
ReFlex Power™
M380056-01 Rev L