Ametek Elgar ReFlex Power Operation Manual page 325

Table of Contents

Advertisement

Active Loads
7. After DC voltage stabilizes, enter the voltage:
8. After DC current stabilizes, enter the current:
9. Send the calibration command to calculate the gain and offset:
10. Send the system error query command to make sure calibration was
11. If there is no error returned at the system error query command,
12. Reset the unit:
7-50
CAL<n>:INP:VOLT 2 <value-2>
CAL<n>:INP:CURR 2 <value-2>
CAL<n>:INP:COND:CALC
successful:
SYST:ERR?
then save the calibration result as follows:
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
ReFlex Power™
M380056-01 Rev L