When an DC power supply module is first powered on, the following
parameters are set to factory default values:
R
ESET
If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:
I
6.5.6
NITIAL
P
OWER
1. Ensure that the AC/DC input service is disconnected.
2. Install the interface cable between the host controller and the
3. Connect an appropriate output load to the DC power supply
6-58
O
PERATIONAL
-O
C
N
ONDITIONS
Output configuration: standalone
Output voltage: 0V
Output current: 0A
Overvoltage protection (OVP): 110% of full-scale voltage;
programmable mode
Undervoltage protection (UVP): 0V
Overcurrent protection (OCP): 120% of full-scale current;
programmable mode
Output state: off
Output isolation and remote sense relays: open
C
ONDITIONS
Output configuration: unchanged
Output voltage: 0V
Output current: 0A
Overvoltage protection (OVP): 110% of full-scale voltage;
programmable mode
Undervoltage protection (UVP): 0V
Overcurrent protection (OCP): 120% of full-scale current;
programmable mode
Output state: off
Output isolation and remote sense relays: open
All status reporting data structures are cleared; enable-masks
are not cleared.