Lpal Cal Pot Calibration - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

ReFlex Power™
LPAL C
7.6.19
This procedure calibrates cal-pot. This calibration is done with the DC
current meter connected between the DC power supply and the Active
Load.
This procedure requires maximum voltage 4.5V and maximum current
15Amps.
1. Connect the shunt between the DC power supply and the active
2. Issue clear and reset commands:
3. Set the DC power supply to 4.5 Volts
4. This begins the CalPots calibration sequence. The Load will be set
5. Program to draw maximum current
6. Adjust the current limit on the DC Power supply to just under the
M380056-01 Rev L
P
C
AL
OT
ALIBRATION
load. Connect the DC current meter across the shunt.
*CLS<n>
*RST<n>
to STATE 1; relays closed and set to 0 current, the 0 current offset
value will be calculated.
CAL<n>:CALPOTS:GAIN:START
CAL<n>:INP:CURR:COUNTS 0xFFFF
limit 13.7 Amps. Use the Current Shunt and Voltmeter to adjust and
measure the current. When the current reading is stable, enter the
current using following command
CAL<n>:CALPOTS:GAIN:CURR <value-1>
Active Loads
7-55