The ReFlex Power™ Controller module should respond with its identification
Verify communications with the ReFlex Power™ AC power supply module by
The AC power supply module should respond with its identification string.
S
TANDALONE
Connect the output for standalone operation per the section, Standalone
Connect appropriate instrumentation for measuring the output voltage and
Perform the power-on verification test per the section, Power-On Check.
Turn on the isolation and remote sense relays, turn on the output converter,
Measure the output voltage and current with the external instruments.
Ensure that the actual values of the output parameters are within
Query the values of the output parameters with the following commands:
Ensure that the readback values of the output parameters are within
Set the frequency to 400 Hz: SOUR[n]:FREQ 400
Close the auxiliary relay: OUTP[n]:AUXR 1
Verify the proper output voltage at the AUX output (should equal 22% of
M380056-01 Rev L
the slot of the ReFlex Power™ Mainframe where the module is located,
counting left to right, 1 through 12.
string.
issuing an SCPI *IDN[n]?, where "n" is a channel number that
corresponds to the slot of the ReFlex Power™ Mainframe where the
module is located, counting left to right, 1 through 12.
The default settings of the module will not be affected.
C
ONFIGURATION
Configuration.
current, as well as a load.
and program full scale voltage and current with the following commands:
*RST[n]
OUTP[n]:STATE 1
SOUR[n]:VOLT <fval>
SOUR[n]:CURR <fval>
specifications of the programmed values.
MEAS[n]:VOLT?
MEAS[n]:CURR?
specifications of the actual values.
VOLT fval as entered in Step 4 above).
O
C
PERATION
HECK
AC Power Supplies
4-41