Hall Calibration For Voltage Range High; All Alibration For Oltage Ange Igh (1.2 Kh Nit ) - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

AC Power Supplies
H
4.6.23
ALL
U
NIT
This procedure calibrates the hall gain sensor of the power supply for the 1.2KHz
AC unit only. This is a seven-point calibration procedure.
1. Close the switch between the power supply and the resistive load.
2. Connect the meter between the output and adjustable resistive load.
3. Set the voltmeter to read the AC current.
4. Ensure that load is connected to the output.
5. Issue clear and reset commands:
6. Set the voltage range of the power supply to high:
7. Turn the unit on:
8. Set the hall calibration data to the default value:
9. Set the output voltage to 120 Volts:
10. Set the output
11. Turn on the resistive load and adjust the load until the output voltage is 230 Volts,
this will ensure that the AC supply is in Constant Current Mode.
12. Set the output frequency to 60Hz:
13. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current:
4-82
C
ALIBRATION FOR
)
*CLS
*RST<n>
SOUR<n>:VOLT:RANGE 1
OUTP<n>:STATE 1
CAL<n>:OUTP:FREQ:CURR:DEFAULT
SOUR<n>:VOLT 240
SOUR<n>:CURR 3.15
SOUR<n>:FREQ 60
CAL<n>:OUTP:FREQ:CURR1 <value-1>
V
R
OLTAGE
ANGE
ReFlex Power™
H
(1.2 KH
IGH
Z
M380056-01 Rev L

Advertisement

Table of Contents
loading

Table of Contents