Ametek Elgar ReFlex Power Operation Manual page 271

Table of Contents

Advertisement

DC Power Supplies
9. Measure the output voltage with the external voltmeter. Enter this as value-2 with
10. Program the output voltage to the third value near 50% of full-scale,
11. Measure the output voltage with the external voltmeter. Enter this as value-3 with
12. Program the output voltage to the fourth value near 72.5% of full-scale,
13. Measure the output voltage with the external voltmeter. Enter this as value-4 with
14. Program the output voltage to the fifth value near 95% of full-scale,
15. Measure the output voltage with the external voltmeter. Enter this as value-5 with
16. Save the calibration results with the following commands:
17. Turn off the output:
6-68
CAL[n]:OUTP:VOLT:COUNTS 14202
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 2 <value->2
count = 48852*0.50+768 = 25194):
CAL[n]:OUTP:VOLT:COUNTS 25194
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 3 <value-3>
(count = 48852*0.725+768 = 36185):
CAL[n]:OUTP:VOLT:COUNTS 36185
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 4 <value-4>
(count = 48852*0.95+768 = 47177):
CAL[n]:OUTP:VOLT:COUNTS 47177
the following command:
CAL[n]:OUTP:VOLT:FIVEPOINT 5 <value-5>
CAL[n]:UNLOCK "6867"
CAL[n]:STORE
CAL[n]:LOCK
(The entered voltage data could be verified with the following query:
CAL[n]:OUTP:VOLT:FIVEPOINT?)