Ametek Elgar ReFlex Power Operation Manual page 165

Table of Contents

Advertisement

AC Power Supplies
14. Set the output frequency to 250Hz
15. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
16. Set the output frequency to 440Hz
17. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
18. Set the output frequency to 630Hz
19. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
20. Set the output frequency to 820Hz
21. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
22. Set the output frequency to 1010Hz
23. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
24. Set the output frequency to 1200Hz
25. Wait for AC current to stabilize. Measure the output current using the AC current
meter. Enter the measured current using following command
26. Perform the voltage calibration using following command
4-80
SOUR<n>:FREQ 250
CAL<n>:OUTP:FREQ:CURR2 <value-2>
SOUR<n>:FREQ 440
CAL<n>:OUTP:FREQ:CURR3 <value-3>
SOUR<n>:FREQ 630
CAL<n>:OUTP:FREQ:CURR4 <value-4>
SOUR<n>:FREQ 820
CAL<n>:OUTP:FREQ:CURR5 <value-5>
SOUR<n>:FREQ 1010
CAL<n>:OUTP:FREQ:CURR6 <value-6>
SOUR<n>:FREQ 1200
CAL<n>:OUTP:FREQ:CURR7 <value-7>
ReFlex Power™
M380056-01 Rev L