AC Power Supplies
C
P
4.6.5
AL
This procedure calibrates cal-pot for the voltage range high. This calibration is
done with the AC current meter connected to the output lead of the AC power
supply.
1. Connect the AC current meter to the output lead of the power supply. Ensure that
no load is connected to the output.
2. Issue clear and reset commands:
3. Set the voltage range of the power supply to high
4. Start the internal calibration of the offset pot
5. Start the calibration of the gain pot
6. Set the load box to approximately 3.25 Amps. Wait for the current to stabilize
then make the AC current measurement using the meter connected to the power
supplies output lead.
7. Finish the gain pot calibration
8. Send the system error query command to make sure calibration was successful.
9. If there is no error returned at the system error query command then save the
calibration result with the following commands
10. Optional Commands:
To query the offset pot set value: CAL<n>:CALPOTS:OFFSET?
4-48
C
OT
ALIBRATION
*CLS
*RST<n>
SOUR<n>:VOLT:RANGE 1
CAL<n>:CALPOTS:OFFSET:CALC
CAL<n>:CALPOTS:GAIN:START
CAL<n>:CALPOTS:GAIN:CURR <value>
CAL<n>:CALPOTS:GAIN:CALC
SYST:ERR?
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
V
R
OLTAGE
ANGE
ReFlex Power™
H
IGH
M380056-01 Rev L