Ametek Elgar ReFlex Power Operation Manual page 354

Table of Contents

Advertisement

ReFlex Power™
7. After DC voltage stabilizes. Enter the voltage using SCPI
8. After DC current stabilizes. Enter the current using SCPI commands
9. Send the calibration command to calculate the gain and offset
10. Send the system error query command to make sure calibration was
11. If there is no error returned at the system error query command then
12. Reset the unit
M380056-01 Rev L
commands
CAL<n>:INP:VOLT 2 <value-2>
CAL<n>:INP:CURR 2 <value-2>
CAL<n>:INP:COND:CALC
successful.
SYST:ERR?
save the calibration result with the following command
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
*RST<n>
Active Loads
7-79