ReFlex Power™
D
7.5.4
EFAULT
P
-O
OWER
N
When an Active Load module is first powered on, the following parameters
are set to factory default values:
R
C
ESET
ONDITIONS
If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:
M380056-01 Rev L
Figure 7-9. Standalone Input Configuration
O
C
PERATIONAL
C
ONDITIONS
Input configuration: standalone
Input current: 0A
Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode
over power protection (OPP): 105% of the Load/s power rating
Overcurrent protection (OCP): 120% of full-scale current; programmable mode
Input state: off
Input isolation and remote sense relays: open
Input configuration: unchanged
Input current: 0A
Overvoltage protection (OVP): 110% of full-scale voltage; programmable mode
Undervoltage protection (UVP): 0V
Overcurrent protection (OCP): 120% of full-scale current; programmable mode
Input state: off
Input isolation and remote sense relays: open (if present)
All status reporting data structures are cleared. Enable-masks are not cleared.
ONDITIONS
Active Loads
7-27