Calibration; Scope - Ametek Elgar ReFlex Power Operation Manual

Table of Contents

Advertisement

ReFlex Power™
13. Ensure that the readback values of the output parameters are
14. To turn off the output and break down the group with the following

CALIBRATION

6.6
S
6.6.1
COPE
Procedures are provided in the following sections for calibration of the ReFlex
Power™ DC power supply modules: RFP-D1016-021-XXXX, RFP-D1065-5A1-
XXXX, RFP-D2033-030-XXXX, RFP-D2050-020-XXXX, RFP-D2050-025-XXXX,
RFP-D2120-8A3-XXXX, and RFP-D2450-2A3-XXXX. Calibration is easily performed
through the Controller module interface with SCPI commands, requiring only a
voltmeter and current shunt.
WARNING
The calibration procedures are performed with the output of the DC power supply
module energized. Do not touch any of the output connections which could be at
hazardous potentials. Calibration must be performed by qualified personnel who
understand the associated hazards and could appropriately contend with them.
CAUTION
If calibration is not performed properly, functional problems could arise, preventing
the DC power supply module from operating properly.
M380056-01 Rev L
MEAS[2000+n11]:VOLT?
MEAS[2000+n11]:CURR?
within specifications of the actual values.
command:
*RST[n]
DC Power Supplies
6-65