the heaviest gauge practical to minimize line drop. The remote sense wires
should be twisted, and may need shielding depending on the electrical noise
environment of the application. Figure 5-8 shows a standalone load connection
with remote sensing.
D
5.5.4
EFAULT
P
-O
OWER
N
When the FPDC power supply module is first powered on, the following
parameters are set to factory default values:
R
C
ESET
ONDITIONS
If a module is reset with the SCPI command, *RST[n], (where n is the module
address), the following default conditions are set:
5-26
(+)
Output
(-)
(+)
Sense
(-)
Power Supply
Figure 5-8. Each Output Voltage Configuration
O
PERATIONAL
C
ONDITIONS
Output configuration: standalone
Output voltage: 0V
Output isolation and remote sense relays: open
Output configuration: unchanged
Output voltage: 0V
Output isolation and remote sense relays: open
All status reporting data structures are cleared. Enable-masks are not
cleared.
Twisted-Pair Cable
Twisted-Pair Cable
(shielded if noise present)
C
ONDITIONS
ReFlex Power™
(+)
Input
(-)
Load
M380056-01 Rev L