Ametek Elgar ReFlex Power Operation Manual page 159

Table of Contents

Advertisement

AC Power Supplies
15. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
16. Set the output frequency to 1000Hz
17. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
18. Set the output frequency to 2000Hz:
19. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
20. Set the output frequency to 3000Hz:
21. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
22. Set the output frequency to 4000Hz:
23. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
24. Set the output frequency to 5000Hz:
25. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
26. Perform the voltage calibration:
27. Send the system error query command to make sure calibration was successful:
4-74
CAL<n>:OUTP:FREQ:CURR2 <value-2>
SOUR<n>:FREQ 1000
CAL<n>:OUTP:FREQ:CURR3 <value-3>
SOUR<n>:FREQ 2000
CAL<n>:OUTP:FREQ:CURR4 <value-4>
SOUR<n>:FREQ 3000
CAL<n>:OUTP:FREQ:CURR5 <value-5>
SOUR<n>:FREQ 4000
CAL<n>:OUTP:FREQ:CURR6 <value-6>
SOUR<n>:FREQ 5000
CAL<n>:OUTP:FREQ:CURR7 <value-7>
CAL<n>:OUTP:FREQ:CALC
SYST:ERR?
ReFlex Power™
M380056-01 Rev L