ReFlex Power™
4. Configure the host controller for communications with the ReFlex
5. Turn on the AC/DC input service.
6. The modules will first execute a power-on self-test. Also the
7. In the following paragraphs, commands include a channel
8. Additionally, model specific parameter values must be added for
9. Verify communications with the ReFlex Power™ Controller
10. The ReFlex Power™ Controller module should respond with its
11. Verify communications with the ReFlex Power™ DC power supply
M380056-01 Rev L
Power™ Controller module.
internal fans are run at maximum speed for several seconds, and
then back to slow speed.
number, [n]. The channel number corresponds to the slot of the
ReFlex Power™ Mainframe where the module is located, counting
left to right, 1 through 12. For example:
SOUR[n]:VOLT 5 will be
SOUR1:VOLT 5 for slot-1
SOUR2:VOLT 5 for slot-2
full-scale voltage and current. They are shown in the following
examples as floating point numerical values, <fval>. For the DC
power supply modules, the values are as follows:
Full-Scale Voltage:
RFP-D1016-021-XXXX: 16
RFP-D1065-5A1-XXXX: 65
RFP-D2033-030-XXXX: 33
RFP-D2450-2A3-XXXXA3: 450
Full-Scale Current:
RFP-D1016-021-XXXX: 20.6
RFP-D1065-5A1-XXXX: 5.1
RFP-D2033-030-XXXX: 30
RFP-D2450-2A3-XXXXA3: 2.3
Therefore, for an RFP-D1016-021-XXXX module in slot 1, the command
for full-scale voltage would be SOUR1:VOLT 16, while for an RFP DC
065-5A1 module it would be SOUR1:VOLT 65.
module by issuing an SCPI *IDN[n]?, where "n" is a channel
number that corresponds to the slot of the ReFlex Power™
Mainframe where the module is located, counting left to right, 1
through 12.
identification string.
module by issuing an SCPI *IDN[n]?, where "n" is a channel
number that corresponds to the slot of the ReFlex Power™
Mainframe where the module is located, counting left to right, 1
through 12.
DC Power Supplies
6-59