AC Power Supplies
14. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
15. Program the output current to 50% of the full scale current:
16. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
17. Program the output current to 72.5% of the full scale current:
18. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
19. Program the output current to 95% of the full scale current:
20. Wait for AC current to stabilize. Measure the output current using the AC meter.
Enter the measured current:
21. Turn the unit off:
22. Perform the current calibration:
23. Send the system error query command to make sure calibration was successful:
24. If there is no error returned at the system error query command then save the
calibration results:
25. Optional Commands:
4-68
CAL<n>:OUTP:CURR:FIVEPOINT2 <value-2>
CAL<n>:OUTP:CURR:COUNTS 2276
CAL<n>:OUTP:CURR:FIVEPOINT3 <value-3>
CAL<n>:OUTP:CURR:COUNTS 3527
CAL<n>:OUTP:CURR:FIVEPOINT4 <value-4>
CAL<n>:OUTP:CURR:COUNTS 4451
CAL<n>:OUTP:CURR:FIVEPOINT5 <value-5>
OUTP<n>:STAT OFF
CAL<n>:OUTP:CURR:CALC
SYST:ERR?
CAL<n>:UNLOCK "6867"
CAL<n>:STORE
CAL<n>:LOCK
ReFlex Power™
M380056-01 Rev L